Directions for Using a Buret
A buret must be scrupulously clean before it is used, and its valve must be liquid-tight.
Cleaning
Thoroughly clean the tube of the buret with detergent and a long brush. Rinse thoroughly with tap water and then with distilled water. Inspect for water breaks. Repeat the treatment if necessary.
Lubricating a Glass Stopcock
Carefully remove all old grease from a glass stopcock and its barrel with a paper towel and dry both parts completely. Lightly grease the stopcock, taking care to avoid the area adjacent to the hole. Insert the stopcock into the barrel and rotate it vigorously with slight inward pressure. A proper amount of lubricant has been used when (1) the area of contact between stopcock and barrel appears nearly transparent, (2) the seal is liquid-tight, and (3) no grease has worked its way into the tip.
Notes
Filling
Make certain the stopcock is closed. Add 5 to 10 mL of the titrant, and carefully rotate the buret to wet the interior completely. Allow the liquid to drain through the tip. Repeat this procedure at least two more times. Then fill the buret well above the zero mark. Free the tip of air bubbles by rapidly rotating the stopcock and permitting small quantities of the titrant to pass. Finally, lower the level of the liquid just to or somewhat below the zero mark. Allow for drainage (, 1 min), and then record the initial volume reading, estimating to the nearest 0.01 mL.
Titration
Figure1 illustrates the preferred method for manipulating a stopcock. When you position your hand as shown, your grip on the stopcock tends to keep the stopcock firmly seated.
Figure 1 Recommended method for manipulating a buret stopcock.
Be sure the tip of the buret is well within the titration flask, and introduce the titrant in increments of about 1 mL. Swirl (or stir) constantly to ensure thorough mixing. Decrease the volume of the increments as the titration progresses, and add titrant drop by drop as you reach the immediate vicinity of the end point (Note 2). When it appears that only a few more drops are needed to reach the end point, rinse the walls of the container (Note 3). Allow the titrant to drain from the inner wall of the buret (at least 30 seconds) at the completion of the titration. Then record the final volume, again to the nearest 0.01 mL. Notes
|
|